To check whether no amplification is caused by kir or instrument;
- Deselect the Subtract Baseline and Normalization functions
- If the lines are unexpected curves as above then it proves that instrument is collection data so you have to focus on your protocol or kits
- If the there is only a single line then it means that instrument is not collection data from the samples.
- Self Test: Go to main menu on the touch screen then; Tools/System Set/Self Test/Start
- Self test takes around 5 minutes and report you the functions of the instruments: OK (Green, Passed) or FAIL (Red)